MECHANICAL VENTILATION DEVICE
Reference number: 20089/2023MECHANICAL VENTILATION DEVICE
1 The device should be intended for operation and ventilation of adult and pediatric patients, namely invasive, non-invasive ventilation and oxygen therapy with respiratory range tidal volume 20 ml. - 2000 ml.
2 Be placed on its own trolley, with lockable wheels.
3 The device should provide propulsion through a built-in own turbine that works independently of a central compressed air system
4 The device should be able to be used for inter-clinical transport of patients and in an ambulance without the addition of an additional compressor.
5 The device should have an emergency valve in case of interruption of oxygen supply, to automatically switch the device to spontaneous breathing mode through filtered ambient air.
6 To have a software procedure for easy and quick aspirating, whereby by pressing a button, the device switches to suction mode, and after completion of the same, an automatic valve is turned on that oxygenates the patient for at least 2 minutes.
7 To have an outlet for connecting to a pneumatic medication nebulizer that is synchronized with inspiration.
8 Built-in independent power supply through an internal battery (minimum battery operation 45 min.) and a minimum of up to 4 hours optionally with an additional battery
9 ventilation modes:
9.1. spnCPAP/PS
9.2. VC-CMV
9.3. VC-SIMV
9.4. VC-AC
9.5. VC-MMV
9.6. PC-APRV
9.7. PC-AC
9.8. PC-BIPAP (Bi-Level)
9.9. NIV – Non-invasive ventilation in all ventilation modes (pressure and volume controlled) with automatic leak compensation
9.10. An advanced option of volume-oriented ventilation modes that delivers the set volume, under the lowest possible pressure or equivalent (PRVC, AutoFlow, ASV or similar)
9.11. O2 therapy using masks for non-invasive ventilation
10 A color screen, at least 12 inches in size, with a touch-sensitive screen.
11 Ability to simultaneously display on the screen 3 curves (pressure, volume, flow) trends.
12 To have measurement of concentration of etCO2 - capnography
13 Fan settings:
- Respiratory (tidal) volume 20ml – 2L
- Inspiratory time 0.2 – 10 seconds (duration of inspiratory phase)
- Frequency of ventilation 2 – 80 per minute
- Inspiratory pressure 1 – 99 mbar
- PEEP from 0 – 50 mbar
- Support for spontaneous breathing of the patient (Plessure support) 0-50 mbar
- Maximum inspiratory flow 250 L/min
- Adjustment of O2 concentration 21 – 100%
- Setting the sensitivity of the trigger 1 – 15 L/min
To have the function of retaining the inspirium for the purpose of x-ray imaging
14 To have leakage compensation on the trigger
15 Automatic tube compensation
16 Display parameters display:
- Pressure (maximum, plateau, medium)
- Total MV, MV for spontaneous breathing
- Tidal volume VT, inspiratory and expiratory VT
- Frequency of breathing (total and spontaneous)
- FiO2
- Inspiratory gas temperature
etCO2
a statement
that in the last 5 years no final court verdict has been issued for a committed criminal offense from Article 88
paragraph 1 of the Civil Code - confirmation that no procedure has been opened for bankruptcy from a competent
authority - confirmation that no liquidation procedure has been opened by a competent authority - confirmation of
paid taxes, contributions and other public duties from a competent authority of the country where the economic
operator is registered - confirmation from the Register of penalties for committed criminal acts of the legal persons
that he has not been sentenced to a secondary penalty, a ban on participation in procedures for public invitations,
awarding public procurement contracts and public-private partnership contracts - confirmation from the Register of
Penalties for Criminal Offenses of Legal Entities that he has not been sentenced to a secondary penalty temporary or
permanent ban on the performance of a separate activity - Certificate that no misdemeanor sanction has been imposed
on him by a final judgment - ban on the performance of a profession, activity or duty, i.e. temporary ban on the
performance of a separate activity
Certificate of registered activity as proof that he is registered as a natural
or legal person for performing the activity related to the subject of purchase or proof that he belongs to an
appropriate professional association in accordance with the regulations of the country where he is registered. -
Solution for wholesale sales of medical devices in accordance with the Law on Medicines and Medical Devices
PHI Clinical Hospital Tetovo, ul. Memet Pash Derala bb Tetovo, Republic Of North Macedonia